OilLab 620 RECC Rapid Equilibrium Closed Cup​

Standards

ASTM D3828, IP 303, EN ISONF3679

 

Marchio CE

 

Subject

These test methods cover procedures for the determination of the flash point by a small scale closed tester. The procedures may be used to determine the actual flash point temperature of a sample or whether a product will or will not flash at a specified temperature (flash/no flash).

 

Measuring R.E.C.C. Principle

A specimen of a sample is introduced by means of a syringe into the cup of the selected apparatus that is set and maintained at the specific temperature/expected flash point. After a specified time a test flame is applied and the observation made whether or not a flash occurred.

 

Measuring R.E.C.C. Devices

  • Testing unit equipped with two ignition systems

  • Electrical system or flame exposure device

  • Flash point detected by a ionization system

 

Measuring Temperature Probe

  • Platinum resistance PT100 class A

 

Measuring Parameters

  • Temperatures: in °C
  • Measuring range: –50°C … +350°C
  • Resolution: 0.06 °C
  • Accuracy: ± 0.1 °C
  • Repeatability / Reproducibility: as per standards methods or better

 

LabLink operating Software

  • User friendly interface

  • All analytical parameters recorded

  • Customizable analysis parameters and methods

  • Customizable results report

  • Printable graphs and results

  • Self-identification of the typology of the analysers connected

 

The software includes:

Analysis Menu

  • Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference

  • Unknow sample

  • Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions

 

Diagnostic Menu

  • Direct access to all analog, digital, inputs and outputs

  • Selectable value displaying: °C / Volt

 

Calibration Menu

  • Automatic calibration of each temperature probe

  • Last calibration date referred to each single probe displayed and relative data printable

  • Display of calibration diagram

  • Insertion of offset values

  • Standard and advanced calibration modes

 

Data Utilities

  • Fields for introduction of operator and product name

  • Archive viewer for files recall

  • All analysis stored in Excel® compatible format

  • Storage capacity for more than 60’000 analysis

  • LIMS compatible

 

Integrated Touch Screen Panel PC

  • TFT/LCD 8"
  • Resolution 1024 × 768, 16.2 M colours
  • USB port for connection to an external printer and/or external PC
  • Storage capacity for more than 60'000 analysis

 

Test Cup

The cup is made of aluminium and equipped with high temperature resistant o-ring. 

 

Automatic opening, closing and positioning of the sliding shutter.

 

The test cup is completely made of aluminium and is equipped with a high temperatures resistant o-ring, allowing an uniform sealing of the closing cover.

 

Particular attention has been paid to the heating system allowing the best heat’s transmission without dispersion on air. The power of the heater is therefore of only 420W but allows to obtain a temperature higher than 370°C.Particular attention has been paid to the heating system allowing the best heat’s transmission without dispersion on air. The power of the heater is therefore of only 420W but allows to obtain a temperature higher than 370°C.

 

The ionisation components are the essential flash point detection system that grants excellent results and high repeatability. The quantity of sample (2 ml / 4 ml) is injected into the cup throught the filling orifice.The instrument is equipped with two ignition systems. An electric pilot that ignites the test flame 30 seconds before the test.

 

Heating

  • Electrical heating cartridges
  • Equipped with over temperature cut-outCooling SystemAir fan

 

Warning light and acoustic signal

When the test temperature is reached, the light blink and an acoustic signal is emitted to inform the operator that the sample must be injected.If the injection of the sample is not performed and confirmed the signal will be repeated after 30 seconds.

 

Shutter

  • Automatic mechanism opening the shutter conform to the methods

 

Electrical Supply

  • 220V ± 15% / 50 to 60 Hz
  • 115V ± 15% / 60 Hz

 

Cord Cable

  • 3 conductors flexible cable with schuko plug

 

Ambient Temperature

  • Max 35°C
  • H.R. 80%

 

Dimensions

  • width 31 cm
  • depth 47 cm
  • height 52 cm

 

Weight

  • 27 kg

 

Spare Parts

  • LAB-620/05-13: heating cartridges

  • LAB-620/06-21: gas valve

  • LAB-620/07-01: electrical ignitor

  • LAB-620/07-03: micro switch

  • LAB-620/07-05: gas ignitor

  • LAB-620/08-12: PT100 sensor

  • LAB-620/08-13: detection / Ionisation cable

  • LAB-620/09-04: gas reducer

  • LAB-620/10-04: PCB fuses, box of 10

  • LAB-620/10-05: main electronic board

  • LAB-620/11-01: silicon tubing, 1 meter

  • LAB-620/12-01: voltage transformer for ignitor

  • LAB-620/13-01: high temperature resistant o-ring for cup

 

Calibration Tools

  • OilLab 80: calibration decade box – PT100 simulator

  • OilLab 81: set of connectors and cables for cold range