OilLab 6000
Standards
ASTM D93 procedures A, B, CDIN EN 22719IP 34ISO 2719
CE marked
Subject
Flash Point on petroleum products, gas oils, fuel oils, lubrificants, biodiesel. Suitable for flash point detection on different substances, waste materials, solvents…
Measuring Pensky Martens Principle
The sample is heated and stirred at specified rates, using one of three defined procedures (A, B, or C). An ignition source is directed into the test cup at regular intervals with simultaneous interruption of the stirring, until a flash is detected.
Measuring Parameters
Temperatures: in °C / °F
Measuring range: 0°C … +420°C
Resolution: 0.01 °C
Accuracy: ± 0.1 °C
Repeatability / Reproducibility: as per standards methods or better
Measuring Temperature Devices
Sample temperature: PT100 sensor completely made in stainless steel resistant to corrosion and shock resistant
Bath temperature: PT100 sensor
Pressure sensor
Built-in sensor with automatic correction of results to a barometric pressure of 101.3 kPa automatically performed by the software at the end of analysis
Detection system
A single multi-detector combines the ionization detector and the thermal detector
Double ignition system
Gas
Ectrical ignitor
Heater
Electrical heating with heating rates as per procedures A, B, C
Stirrer
An electric motor drives the stirring of the product
Stirring speed as per selected procedures A, B, C
Cooling system
Built-in forced air fan
Safety Devices
Automatic diagnostic in case of breakage of the sample temperature probe and thermal sensors
Automatic fire detection system
Overheating protection with auto shut off during the test
Gas valve for closing the gas supply (max 30 mBar), at the end of the test
Fire Extinguisher
Automatic fire detection and extinguisher system
Software Features
All analytical parameters recorded
Customizable analysis parameters and methods
Customizable results report
Printable graphs and results
The software includes:
Analysis Menu
Automatic execution of the analysis in accordance to the selected procedure (Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference as well as costumized procedures)
Automatic handling of samples with unknown flash point
Display in real time of all the analysis parameters and status
Fields for introduction of operator and product name
Expected flash point temperature programmable
Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
Configuration menu with up to 20 preset samples and expected flash point
Automatic barometric correction of results
Diagnostic Menu
Direct access to all analog, digital, inputs and outputs
Selectable value displaying: °C / Volt
Calibration Menu
Automatic calibration of each temperature probe
Up to 100 calibration points (standard with 5 and dynamic with up to 100 points)
Programmable calibration frequency with selectable validity period and notice/lock-down at expiration date
Last calibration date referred to each single probe displayed and relevant data printable
Display of calibration diagram
Data Utilities
Archive viewer for files recall
All analysis stored in Excel® compatible format
LIMS compatible
Access via password to the following menu:
calibration
program settings
selection ignitor type
Calibration:
comparative with reference thermometer and barometer
selection calibration due date
automatic diagnostic in case of faulty of the temperature and thermal sensors
Analysis storage:
storage capacity for more than 65.000 analysis
export of test results files in the most common formats Excel and PDF
Lin-Tech algorithm temperature ramp and auto-adapt for any type of sample
Double Flash detection (by thermal sensor and ionization ring simultaneous)
Rapid test selection between electrical ignitor and gas ignitor
Automatic barometric correction of the results via a built-in barometer
LIMS communication capable
Reading interval of PT100 sample from 0 to 450 °C with resolution 0.01 °C
Reading interval of PT100 bath from 0 to 450 °C with resolution 0.01 °C
Double ignition system: gas and / or electrical ignitor
Fully automatic arm that can be brought into three positions: cleaning, testing and idling time
Integrated Touch Screen Panel PC
TFT/LCD 8.4"
Resolution 1024 × 768, 16.2 M colours
2 × USB ports for connection to an external printer
1 × ethernet port for LAN and LIMS network
Electrical Supply
220V ± 15% / 50 to 60 Hz
115V ± 15% / 60 Hz
Ambient Temperature
Max 35°C
H.R. 80%
Dimensions
width 37 cm
depth 52 cm
height 32 cm
Weight
30 Kg
Accessories / Spare Parts
LAB-6000/05-23: heater collar
LAB-6000/06-21: gas valve
LAB-6000/07-01: electrical ignitor
LAB-6000/07-03: micro switch
LAB-6000/07-04: handle
LAB-6000/07-05: gas ignitor
LAB-6000/08-12: PT100 product
LAB-6000/08-13: detection / ionisation cable
LAB-6000/08-14: PT100 Bath
LAB-6000/09-04: gas reducer
LAB-6000/09-05: calibrated brass crucible
LAB-6000/10-04: PCB fuses, box of 10
LAB-6000/11-02: stirrer belt
LAB-6000-118: fire extinguisher system
Calibration Tools
OilLab 80: calibration decade box – PT100 simulator
OilLab 91: set of connectors and cables for OilLab 6000
Alternative Products
A sentence here about related products with image links below to products you want customer to consider.
SUBMIT
Name:
This field is required.
Email:
This field is required.
Phone:
This field is required.
Industry:
This field is required.
Company Name:
This field is required.
Address:
This field is required.
Post Code:
Country:
This fi.