ERAFLASH
Standards
ASTM D7094, ASTM D6450
Correlation to
ASTM D56, ASTM D93, ASTM D3828, DIN 51758, ISO 2719, ISO 3679, ISO 3680, ISO 13736, IP34, IP170, JIS K2265
THE SAFE SIDE OF FLASH POINT TESTING
The ERAFLASH measures according the continuously closed cup flash point testing methods ASTM D7094 and ASTM D6450. ASTM D7094 shows no statistical bias to the former ASTM D93 Pensky Martens method. The sample volume is only 2 ml. During measurements an electric arc ignites the sample vapor and the analyzer determines the flash point by the pressure change inside the closed cup. Flash point testing never was safer and easier.
ASTM D7094 and ASTM D6450
ASTM D6450 was the first Continuously Closed Cup Flash Point (CCCFP) and uses 1 ml sample volume only. However, the obtained results didn’t compare well to the well-established ASTM D93 Pensky Martens method. Therefore a modified successor method was developed.
ASTM D7094 (Modified CCCFP) uses 2 ml of sample volume and a different heating rate than ASTM D6450 with the effect that it shows no statistical bias to ASTM D93 (Pensky Martens). This was proven by an ASTM round robin test in 2004.
In addition this round robin showed, that the actual repeatability and reproducibility using Pensky Martens flash point testers is up to 50 % worse than the values stated in ASTM D93.
Fuel Specifications Testing
In December 2013 ASTM D7094 was accepted as the first official alternative for fuel specification testing as it gives equivalent results to ASTM D93 and even outperforms it in terms of reproducibility:
- Fuel oil (ASTM D396)
- Diesel fuel (ASTM D975)
- Gas turbine fuel oil (ASTM D2880)
- Kerosene (ASTM D3699)
- Diesel Fuel, Biodiesel Blend (ASTM D7467)
Unmatched Range Meets High Performance
eralytics’ patented PBT – Peltier Boost Technology™ allows flash point testing below 0 °C (32 °F) and above 200 °C (392 °F) with a single instrument. This technology separates the Peltier elements during the heat up from the oven to prevent any damage to them at high temperatures. Additionally the Peltier elements can be used during cool down at temperatures that would normally harm them. Thus, the turnaround time is significantly shortened. On top PBT makes ERAFLASH the only flash point tester on the market covering a temperature range of -25 °C (-13 °F) to 420 °C (788 °F).
With its innovative design ERAVAP is capable of measuring gasoline, oxygenated gasoline as well as any other solvent or even polymers. It is used worldwide by refineries, terminals, oil and pipeline companies as well as authorities and independent laboratorie
High Sample Throughput with Minimum Maintenance
The innovative PBT – Peltier Boost Technology™ allows previously unmatched heating and cooling rates, which in combination with the low sample volume decreased the turnaround time significantly for flash point testing. Additionally the CPT – Contamination Prevention Technology™, which is a combination of advanced electrode protection and self-cleaning procedures, reduces the time required for cleaning and maintenance to an absolute minimum, increasing the sample throughput even further.
.
Measurement Procedure
Just four easy steps are required for a safe flash point testing at maximum precision:
Enter sample ID and operator name
Select the measurement standard and add additional parameters (e.g. expected flash point)
Fill the sample cup with sample, place the sample cup inside the measurement chamber, and close the door
Press the RUN button
The instrument now visualizes the measurement procedure on the screen showing the sample and oven temperature as well as pressure gauge displaying the pressure increase after ignition. After the measurement the flash point temperature is corrected for the atmospheric pressure and rounded to the next 0.5 °C. The result will be saved in the corresponding result file.
Characteristics of the flash point testing can be seen in the Combustion Graphics™. This makes ERAFLASH the ideal solution for R&D as well as quality control.
Alternative Products
A sentence here about related products with image links below to products you want customer to consider.
Feature Video
- Safe flash point testing with no open flames or glowing wires
- Highest temperature range on the market with a single analyzer
- Unmatched heating and cooling rates using PBT- Peltier Boost Technology™
- Minimum sample volume for expensive samples
- Easy usage and highest performance
- One analyzer for all applications
Operation Video
Watch this video to get an overview on the sample handling and measurement procedure of ERAFLASH.
SUBMIT
Name:
This field is required.
Email:
This field is required.
Phone:
This field is required.
Industry:
This field is required.
Company Name:
This field is required.
Address:
This field is required.
Post Code:
Country:
This fi.
ERAFLASH
For more information on this product or to arrange a call from one of the Cea Petrochemical Team, please complete the contact form below.