PAMAS S50P Fuel
PAMAS S50P Fuel
The online particle counter PAMAS S50P Fuel is used as fixed installed measuring instrument for condition monitoring and contamination analysis of liquid fuel. The instruments measures the particle number in eight different size channels and and early alerts in case of failures caused by bigger particle sizes (e.g. corrosion). Due to this feature, the particle analysing system PAMAS S50P Fuel is indispensable for preventive condition monitoring of liquid fuel, as it prevents the filling of contaminated fuel into the tank. On its 6-digit LED display, the instrument reports the triple code according to ISO 4406. The PAMAS S50P Fuel is equipped with a pump controlling the flow rate to 25 ml/min at a pressure range from 0 to 6 bar.
Application
Contamination control and condition monitoring of liquid fuel
Sample input
Pressure range: 0 to 6 bar
Sample flow rate: 25 ml/min
Volumetric particle sensor PAMAS HX
Calibration range: 4 to 70 µm(c) as per calibration standard ISO 11171
Measurement in eight size channels: > 4 µm(c), > 6 µm(c), > 10 µm(c), > 14 µm(c), > 21 µm(c), > 25 µm(c), > 38 µm(c) and > 70 µm(c)
Maximum particle concentration: 24.000 particles per ml at a coincidence quote of 7.8% and a flow rate of 25 ml/min
Standardised measuring results
Report of measuring results according to cleanliness class standards SAE AS 4059 and ISO 4406.
The particle counter measures ISO classes from 0 to 22.
Interfaces for data transfer
Standard equipment: RS485
Optional equipment: analogue 4-20 mA interface. Parallel data transmission for the size channels > 4, > 6, > 14 and > 70 µm(c) or serial data transmission for all eight size channels.
Software
PAMAS POV (PAMAS Online Visualisation): Software for online visualisation of measuring results and for long-term trend monitoring
Alternative Products
A sentence here about related products with image links below to products you want customer to consider.
PAMAS S50P Fuel
For more information on this product or to arrange a call from one of the Cea Petrochemical Team, please complete the contact form below.
SUBMIT
Name:
This field is required.
Email:
This field is required.
Phone:
This field is required.
Industry:
This field is required.
Company Name:
This field is required.
Address:
This field is required.
Post Code:
Country:
This fi.